奥林巴斯Olympus 5058PR脉冲发生器/前置放大器奥林巴斯5058PR脉冲放大器5058PR脉冲发生器接收器

5058PR脉冲发生器接收器/北京泰如专业供应平台,5058PR前置放大器/欢迎来电索取价格及产品资料010-57464568

奥林巴斯Olympus 5058PR脉冲发生器接收器/北京泰如电子科技价格优惠,奥林巴斯各种规格扫查架专业供应



本着为“客户提供**的产品和**惠的价格以及**的售后服务”以宗旨,北京泰如电子科技有限公司在保证产品质量、性能的同时,还会定期回访,进行仪器与配件保养、维护及技术交流。让客户能得到**的服务、满意的价格、放心的使用。

5058PR高压超声脉冲发生器接收器专用于对声束衰减性强的材料的检测和测量应用。其脉冲发生器**高可发射900伏激励脉冲,从而可适当对低频探头进行补偿。

The Panametrics Model 5058PR High Voltage pulser-receiver is designed especially for ultrasonic test and measurement applications in attenuating mediums. The pulser section of the Model 5058PR can deliver up to 900 volts in an impulse type excitation pulse to appropriate low frequency transducers. The flexible receiver section provides switchable 40/60 dB gain with an additional 30 dB available from an integral low noise preamplifier. Four position high and low pass filters and 1 dB attenuation steps further optimize signal response.

A full range of front panel controls permits discrete calibrated settings or continuous adjustments for easy and repeatable instrument setup. The unit is packaged in a rack adaptable 19" (483 mm) wide enclosure.

Features & Benefits
• High voltage pulser, switched settings or continuously adjustable to 900 volts
• High gain, low noise broadband (10 MHz) receiver
• Pulse-echo and thru-transmission modes
• High isolation (80 dB) in thru-transmission mode
• 60 dB RF Gain
• Additional 30 dB gain available from internal auxiliary preamplifier
• Receiver attenuation range of 0 to 80 dB in 1 dB steps
• 6 dB vernier for fine adjustment
• Switchable high pass and low pass filters
• PRF rate switch selectable from 20 Hz to 2 kHz
• Controls permit either discrete, calibrated settings or continuous adjustment

Applications
Used with an analog or digital oscilloscope, the Model 5058PR can be the basis of a bench top measurement system for determination of material quality or properties. Flaw detection, time of flight, thickness, velocity, attenuation, and frequency content are investigated. The high voltage pulser and high gain receiver of the Model 5058PR make it particularly useful for applications that challenge the penetrating capability of conventional pulser/receivers in such materials as:
- coarse grain cast metals
- concrete, stone, soil or sand
- composites and fiberglass
- very thick plastics or rubber
- highly attenuating liquids, slurries or emulsions
- wood and fiberboard
- air gaps

Potential increase in sensitivity using a tuned square wave pulser versus a spike pulser as a function of transducer center frequency.
Response using a Model 5073PR 75 MHz bandwidth pulser-receiver with a V215-BA-RM 50 MHz permanent delay contact transducer coupled to a 0.075 mm (0.003") metal shim
Spike excitation features sharp rise times adjusted by energy and damping values.
Square wave has controlled rise and fall times with directly adjustable voltage and pulse width.
A-scan comparison between 5073PR and 5072PR with 50 ohm damping and energy Position 1 to 5077PR set at 100 volts and tuned for frequency optimization. Gain set to noted dB level to normalize signal height from a 20 MHz delay line transducer.

Specifications

PULSER 5058PR
Pulse Type (Main Bang):
0 to 20 KHz by external trigger pulse
Available Pulse Voltage:
-100 to -900 V continuously variable; -100, -200, -400 or -900 V switch selectable (meter measures available voltage on pulse storage capacitor)
Pulse Rise Time:
<40 nS @ 900 V, <10 n0S @ 100 V
Available Pulse Energy:
11 to 890 microjoules, stepless, or 11, 44, 170, or 890 microjoules in selected pulse voltage positions
Damping:
50 to 333 ohms, continuosly variable; 50, 100, 200, or 500 ohms switch selectable
Pulse Repetition Rate:
Internal: 20, 50, 100, 200, 500, 1 k or 2 kHz, switch selectable 
External: 0 to 2 kHz maximum
Sync Output Pulse:
TTL compatible, or +1v minimum into 50 ohms (occurs approx. 75hS before main bang when externally triggered).
External Trigger Input:
Edge triggered, accepts + or - input pulses. 
Zin = 50 or 500 ohms (internal switch) 
Max pulse height in: 65 V 
Rise time: <30 µS 
Pulse width: 0.1 µS <PW<4 µS 
Delay after Ext. Trig.: <200 nS
Mode:
Pulse-echo or thru-transmission, switch selectable
RECEIVER
Maximum Receiver Input Power:
The maximum average power that receiver input can dissipate is 1W. Sine wave RF input should be limited to 10 V rms or 40 V rms at 25% duty cycle.
Input Impedance:
500 ohms (linear range). Approx. 140 ohms for signals> linear range
Voltage Gain:
40 dB or 60 dB (RL = 50 ohms), switch selectable
Phase:
Normal (0°) or inverted (180°), switch selectable
Bandwidth:
10 kHz to 10 MHz (-3 dB)
High Pass Filters:
0.01 (out), 0.03, 0.1, 0.3 or 1.0 MHz, switch selectable
Low Pass Filters:
0.5, 1.0, 3.0, 5.0, or 10 MHz (out), switch selectable
Attenuation:
0 dB to 80 dB in 1 dB steps
Attenuation Vernier:
61 dB continuous (vernier can be switched off) with status LED.
Noise:
Typically 50 µV pk-pk referred to the input
Output Impedance:
50 ohms in series with 10 µF
Maximum Signal Out:
±0.5 V peak (RL =50 ohms), 
±1 V peak (RL ≥ 1 K ohms)
AUXILIARY PREAMPLIFIER
Input Impedance:
100K ohm in parallel with 50 pF.
Maximum Input Power:
The maximum power that the receiver can dissipate is 1 W.
Gain:
30 dB, non-inverting (RL = 500 ohms)
Bandwidth:
600 Hz to 5 MHz (-3dB)
Noise:
25 µV pk-pk referred to the input (Rs = 500 ohms, RL = 500 ohms)
Output Impedance:
50 ohms in series with 4.7 µF
Maximum Output Level:
+/-0.5 V pk (RL = 500 ohms)
UNIT
Signal Connectors:
BNC
Power Requirements:
100, 120, 220, or 240 VAC selectable, 50-60 Hz, 45 watts maximum
Fusing:
0.5A Slow Blow for 100 or 120 VAC, 0.25A Slow Blow for 220 or 240 VAC
Size:
3.5" H x 16.7" W x 12.7" D 
88.9 mm H x 424 mm W x 323 mm D
Weight:
13.3 lbs (6 kg)
INTERIOR ConTROLS (USER ADJUSTABLE)
Ext Trigger Termination:
50 or 500 ohms, selectable (factory set to 50 ohms).
Ext Trigger threshold:
0.1 to 1 V (factory set to 0.5 V)

Standard Inclusions
The Model 5058PR is shipped with two BCB-58-4 cables, one 50 ohm terminator, application detachable power/mains cord, and operating manual. An optional rack-mount kit is available and should be specified at time of ordering.

*All specifications are subject to change without notice.























































































北京泰如电子科技以技术为先导,为客户提供高效、**和安全的检测解决方案。我们设计、并维护超声波探伤仪,根据客户需求定制特殊规格探头,满足于各行业,代理工业工业内窥镜,X射线和涡流检测设备,为航空航天、电力、石油天然汽车和钢铁等行业应用制定专业检查方案。主要代理品牌:美国GE、德国KK、美国奥林巴斯(Olympus)、汕头汕超、美国泰克(Tektronix)、美国福禄克(FLUKE)、美国安捷伦(Agilent)、**固纬(GW)、香港优利德(UNI-T)、新美达(美柯达)、沪船(船牌)等。
主要经营:超声波探伤仪、无损检测仪器系列、示波器系列、安规测试仪系列、分析仪器系列、红外测温仪系列、行业专用仪器系列、电子测试仪器系列、环境监测仪器系列、力学仪器系列、电工测试仪器系列、通讯仪器仪表系列、通用监测仪器系列、工程监测仪器系列、计量检测仪器系列、气体检测仪器系列、其他测试仪器等。

奥林巴斯Olympus工业扫查架一览表;


联系方式
公司:北京泰如电子科技有限公司
状态:离线 发送信件
姓名:张文庆(先生)
电话:010-57464568
手机:13699112842
地区:北京-通州区
地址:北京市通州区土桥中街140号金隅自由筑1-1-1706
邮编:100000
QQ:3081402071
商铺:http://m.ceoie.com/bjtrdz/
去北京泰如电子科技有限公司怎么走?上图中的红点是北京泰如电子科技有限公司在通州区的具体位置标注,操作左上角地图工具可以放大缩小哦。
相关信息
[试验机] 推荐供应
最新发布信息
点击分享到微信、朋友圈、QQ...
字母索引:  A  B  C  D  E  F  G  H  I  J  K  L  M  N  O  P  Q  R  S  T  U  V  W  X  Y  Z

首页 | 公司 | 求购 | 供应 | 商城 | 招商 | 展会 | 行情 | 品牌 | 产品

联系我们 | 刷新 | 返回顶部

©2004-2024  贸易商务资源网(ceoie.com)  版权所有